User guide

page 37 of 61
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999248 / 28/04/2011
Conguration
Adjustment 1013 mbar
RS-232...
Sensors...
Display...
Autostart Off
Defaults Cancel
--------Back--------
Inthemenu”Conguration”thedeviceparametersarepreselected.
Adjustment:Adjustmentofthepressuretransducerundervacuumand/oratatmosphericpressure,
see also section ”Readjustment the controller”.
Adjustment at atmospheric pressure is carried out at an absolute pressure value between 1060 - 700
mbar and under vacuum at an absolute vacuum value between 0 - 20 mbar. In the range between 20
and 700 mbar no adjustment is possible; ---- mbar is displayed.
RS-232: Conguration of the interface, setting of parameters and commands,seesection”Inter-
face”.
Baudratecanbesetto19200,9600,4800or2400,parityon”8-N-1”,”7-O-1”or”7-E-1”,Handshake
on ”no”, ”Xon-Xoff” or ”RTS-CTS” and remote on ”On” or ”Off”.
Sensors: Selectionofthepressuretransducertobecontrolled(maximum4externalsensors).
Display: Selection of the device parameters Brightnessbetween 0 - 100%, Contrastbetween
0 - 100%, Sound”On” or ”Off”, Units ”mbar”, ”hPa” or ”Torr”, Language”German”, ”English”,
”French”, ”Italian”, ”Spanish”, ”Turkish”, ”Korean”, ”Chinese”, ”Portuguese”, ”Russian”, ”Polish”,
”Dutch”,”Japanese”,“Finnish”.
Autostart: If ”Autostart” is set to ”On” the controller restarts a running process automatically after a
mains failure. If this is unwanted, set Autostart to ”Off”.
Attention: If Autostartis preselected, the process starts immediately after power failure without
pressinganyfurtherkey.Itistheuser’sresponsibilitytoensurethatnodangerousstatusofthesys-
tem due to the automatic start-up can occur and to provide appropriate safety measures. If neces-
sary,theuserhastocheckprior to starting the process if the option ”Autostart” is enabled.
Defaults: If Defaultsis set to ”Load”, the controller is reset to factory set values. All stored programs
and parameters are deleted.
The screen-shot shows the factory-set values.
Preselections
Usetheselectionknobtoselecttheparameters.
5.8 Function Conguration