User guide

page 35 of 61
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999248 / 28/04/2011
Select function Pump down”. Speed”: ”HI” is recommended. For gel dryers set a lower speed if
necessaryifthegelstendtobreak.
Set ”Minimum” to prevent volatile components to evaporate. The process is stopped automatically as
soon as ”Minimum” is reached.
Set a process time (”Duration”) if necessary.
Startprocessbypressingkey”START/STOP”.
alternatively:
Select function ”Vac control” to dry at a predetermined vacuum level.
Set ”Set vacuum” to the evaporation vacuum of the solvent.
Set a process time (”Duration”) if necessary.
Startprocessbypressingkey”START/STOP”.
5.7 Examples for use
Assembly of a vacuum system
Assemble vacuum connection lines between controller, vacuum pump and apparatus.
Assemble electrical connections.
Connect coolant if necessary.
5.7.1 Vacuum for ltration and suction
Select function ”Pump down”.
If necessary set value for ”Speed” (high or low pumping speed). Reduce the speed in case of fast
sedimentation(formationofaltercake).
Set ”Minimum” to a value which still does not lead to evaporation of the solvent.
Startprocessbypressingkey”START/STOP”.
alternatively:
Select function ”Vac control”.
If necessary set value for ”Speed” (high or low pumping speed). Reduce the speed in case of fast
sedimentation(formationofaltercake).
Set ”Maximum”toswitchoffthepumpattheendoftheprocessorincaseofaltercakecrack(pres-
sure increase).
Startprocessbypressingkey”START/STOP”.
Tip for ltration: Adjust preset pressure to a value well above the boiling pressure of the solvent
(e.g.water>>20mbar).Setthemaximumpressuretoe.g.500mbar.Oncetheltrationhasnished,
the pressure increases and the pump is switched off automatically.
5.7.2 Vacuum for gel dryer, drying chambers and vacuum concentrators