Preselections
☞ Usetheselectionknobtosettheparameters.
☞ Sensitivity: The ”Sensitivity” of the control effects the control speed. High sensitivity leads to a reduced
pumping down speed, e. g. for small amounts of solvents or foaming processes. Use a low sensitivity for
”uncritical” processes to reduce process time. Usually setting the ”Sensitivity” to ”normal” is appropri-
ate.
The ”Sensitivity” is adjustable to ”high”, ”normal” or ”low”.
☞ Speed:Themaximummotorspeedcanbepreselected(tolimitthepumpingspeed).Theselection
”HI” (recommended) causes an automatic limitation of the adaptation depending on parameters de-
termined during the process. Only in case of very sensitive processes it is advisable to reduce the
maximumspeed.
The ”Speed” is adjustable in a range of 1-100% or can be set to ”HI”.
☞ Minimum: If the vacuum value reaches ”Minimum” the process is stopped. Use ”Minimum” for auto-
matic process termination. When selecting ”Auto” the controller switches off automatically at the end
of the process (boiling plateau), recommended only for large amounts of solvents.
The ”Minimum” is adjustable in a range of 2 - 1060 mbar, or can be set to ”Auto” or to ”Off”.
☞ Delay: ”Delay”speciesthetimethecoolantvalvecontinuestorunafterprocessstop.
The ”Delay” is adjustable in a range of 1-300 minutes or can be set to ”Off”.
☞ Duration: ”Duration” determines the total process time since control start.
The ”Duration” is adjustable between 1-1440 minutes (24 h) or can be set to ”Off”. A preset ”Duration”
(process time) is without effect, if the process is stopped earlier by reaching the preset ”Minimum”.
➨ Control of a pump in function auto mode: Automatic determination of the boiling vacuum and
automatic adaptation of the boiling vacuum in case of changing process parameters.
Auto mode
Sensitivity normal
Speed HI
Minimum Off
Delay Off
Duration Off
- - - - - - Graphic - - - - - -
-------Back-------
00:00:00
The screen-shot shows the factory-set values.
When ”Graphic” is selected the display shows a pressure vs. time curve (right screen-shot).
Temporary switching from ”Auto mode” to function ”Vac control”
☞ Presskey”MODE”.Thecontrollerswitchesto”Vaccontrol”.Thecurrentpressurevalueisusedand
held as new set point.
☞ The preset function of the controller does not change due to this temporary switching. When pressing
key”STOP”thecontrollerissetagaintothefunction”Automode”.
The timeline in the diagram adapts automatically to the process time.
☞ Presstheselectionknobtwicetoreturntothestandarddisplay.
50 mbar
00:07:45
5.4 Function Auto mode
Auto mode