4 Use and operation
4.2 Installing in a vacuum system
• Connectionlinesatthepumpinlethavetobegastight.Particlesanddustmust
notbeaspirated,theuserhastoprovideappropriateltersifnecessary.Theuser
mustensuretheirsuitabilityconcerninggasow,chemicalresistanceandsafe-
ness against clogging prior to use.
• Connectanexhaustlinegastightatthepumpoutletifnecessary.Alwaysdispose
ofexhaustgasesappropriately(e.g.intoafumehood).Ifthereisriskofreleaseof
dangerousorpollutinguids,installanappropriatesystemtocatchanddispose
ofthoseuids.
• Reducethetransmissionofvibrationandpreventloadingduetorigidpipelines.
Insertelastichosesorexibleelementsascouplingsbetweenthepumpandrigid
pipes. Attention:Flexibleelementstendtoshrinkwhenevacuated.
• Theexhaustlinehasalwaystobefree(pressureless)toensureanunimpeded
discharge of gas.
• Especiallyifthegasballastvalveisopen,apowerfailuremaycauseuninten-
tional ventilation of the pump and the vacuum system. In case this constitutes a
potentialsourceofdanger,takeappropriatesafetymeasures.
• Makesureventilationisadequateifthepumpingunitisinstalledinahousingor
ifambienttemperatureiselevated.Provideexternalventingifnecessary.Keepa
distance of minimum 20cm between fan and ambient parts.
Whenswitchingonthecontrollerfortheveryrsttime,amenutoselectthelanguageofthecontroller
menu is displayed. Select the desired language (e.g. ”English”)byturningtheselectionknobandpress
toconrm.Thenselectthepressureunit(”mbar”, ”Torr” or ”hPa”) in the same way.
It is possible to access the language selection menu at any time by switching on the controller
while keeping the selection knob pressed.
4.1 Using a ROTACOOL reux cooler
If using the ROTAVAC vario Pumping unit together with
theROTACOOLreuxcooler,placetheROTAVACvario
Pumping unit directly on top of the ROTACOOL.
If using the ROTAVAC vario Pumping unit without the
ROTACOOL reux cooler, an extra foot has to be in-
stalled for operation with the condensate cooler (14; in-
cluded in the scope of delivery of the condensate cool-
er). Position the ROTAVAC vario Pumping unit simply
ontheextrafoot,therebyplacingthepump’sfeetinthe
fourhollowsoftheextrafoot.
Install the coolant hoses in such a way that the stability
of the pump is assured even with a full catchpot