User guide
page 10 of 61
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 999248 / 28/04/2011
• Complywithapplicableregulationswhendisposingofchemicals.Takeintocon-
sideration that chemicals may be polluted.
Takeadequateprecautionstoprotectpeoplefromtheeffectsofdangeroussub-
stances(chemicals,thermaldecompositionproductsofuoroelastomers),wear
appropriate safety-clothing and safety glasses.
• Useonlygenuine spare parts and accessories. Otherwise safety and perform-
ance of the equipment as well as the electromagnetic compatibility of the equip-
ment might be reduced.
PossiblytheCEmarkortheC/USconformitybecomesvoidifnotusinggenuine
spare parts.
• Switchingofanin-linevalveoropeningofaventingvalvebythecontrollermust
not lead to a critical dangerous situation under any circumstances.
• Attention: If Auto-Start is preselected, the process starts immediately after a
power failure without pressing any further key. It is the user’s responsibility to
ensure that no dangerous status of the system due to the automatic start-up can
occur and to provide appropriate safety measures. If necessary, the user has to
checkprior to starting process control whether the option ”Auto-Start” is ena-
bled.
Donot startthe pumpif thepressuredifferencebetweeninletandoutletport ex-
ceeds1.1baratmaximum.
Preventanybackpressureofgasesandthebackowofcondensates.
Neversuckliquidsordustintothepump.
Provide appropriate protective measures (i.e precautions which allow for the require-
ments of the respective application) even for the case of failure and malfunction.
Failure of the pump (e.g. due to power failure) or of connected components, of parts
of the supply or change of parameters must not lead to a critical dangerous situa-
tionunderanycircumstances.Incaseofdiaphragmcracksorleaksinthemanifold
pumped substances might be released into the environment or into the pump hous-
ing or motor. Comply especially with notes on operation and use and maintenance.
Due to the residual leak rate of the equipment,theremightbeanexchangeofgas,
albeitextremelyslight,betweentheenvironmentandthevacuumsystem.
Adopt suitable measures to prevent contamination of the pumped substances or the
environment.
The controller is powered by a short circuit proof wide-range power supply with
integrated overload protection.
The motor is protected by a temperature sensor at the circuit board: Current
limitation if the temperature at the circuit board is higher than 70°C, switching off the
pumpifthetemperatureishigherthan85°C.Incaseofblockadeofthemotor(after
10 attempts to start-up) the pump is switched off.
If the pump is switched off due to safety measures, manual reset is necessary. Isolate
the pump from mains. Eliminate the cause of failure before restarting the pump.
NOTICE