page 29 of 34
Overpressure safety relief device ........................................
.......................................................................23-30-01-04-98
7.4 Replacing the overpressure safety relief device at the condenser
Round bottom ask 500 ml, coated .............. 514-83000-02
➨ Remove joint clip at the catchpot.
➨ UnscrewthefourTorxscrewsatthecounterholderofthe
condenser and remove condenser. Thereby remove the
adapter from the inlet of the condenser.
➨ Pull off old overpressure safety relief device and install
new one. Check for correct position of the PTFE-foil under
the overpressure safety relief device.
➨ Reassembly in reverse order.
➨ Position adapter in the inlet of the condenser and screw
condenserwithcounterholdertothepump(Torxscrews).
Fixwithunionnut.