4.1 Connection of the control cable
Factory-set, the pump is designed for operation with a control signal; i.e. without a
control signal, the pump does not start!
Connectthecontrolcableofthepumptothevacuumbox(seeinstructionmanual
Hei-VAP Precision).
4.2 Using a ROTACOOL reux cooler
4 Use and operation
4.3 Installing in a vacuum system
• Connectionlinesatthepumpinlethavetobegastight.Particlesanddustmust
notbeaspirated,theuserhastoprovideappropriateltersifnecessary.Theuser
mustensuretheirsuitabilityconcerninggasow,chemicalresistanceandsafe-
ness against clogging prior to use.
• Connectanexhaustlinegastightatthepumpoutletifnecessary.Alwaysdispose
ofexhaustgasesappropriately(e.g.intoafumehood).Ifthereisriskofreleaseof
dangerousorpollutinguids,installanappropriatesystemtocatchanddispose
ofthoseuids.
• Reducethetransmissionofvibrationandpreventloadingduetorigidpipelines.
Insertelastichosesorexibleelementsascouplingsbetweenthepumpandrigid
pipes. Attention:Flexibleelementstendtoshrinkwhenevacuated.
• Assembletheglasssetstress-free.
• Theexhaustlinehasalwaystobefree(pressureless)toensureanunimpeded
discharge of gas.
• Especiallyifthegasballastvalveisopen,apowerfailuremaycauseuninten-
tional ventilation of the pump and the vacuum system. In case this constitutes a
potential source of danger, take appropriate safety measures.
• Makesureventilationisadequateifthepumpingunitisinstalledinahousingor
ifambienttemperatureiselevated.Provideexternalventingifnecessary.Keepa
distance of minimum 20cm between fan and ambient parts.
• Beforeswitchingoncheckthatmainsvoltageandcurrentconformwiththeequip-
ment (see rating plate).
If using the ROTAVAC vario control or Rotavac vario
tectogetherwiththeROTACOOLreuxcooler,place
the ROTAVAC vario control or ROTAVAC vario tec di-
rectly on top of the ROTACOOL.
If using the ROTAVAC vario control or ROTAVAC vario
tecwithouttheROTACOOLreuxcooler,anextrafoot
has to be installed for operation with the condensate
cooler (13; included in the scope of delivery of the con-
densate cooler). Position the ROTAVAC vario control
or the ROTAVAC vario tec simply on the extra foot,
thereby placing the pump’s feet in the four hollows of
theextrafoot.
Install the coolant hoses in such a way that the stability
of the pump is assured even with a full catchpot