4.1 Connection of the control cable
Factory-set, the pump is designed for operation with a control signal; i.e. without a
control signal, the pump does not start!
Connectthecontrolcableofthepumptothevacuumbox(seeinstructionmanual
Hei-VAP Precision).
4.2 Using a ROTACOOL reux cooler
4 Use and operation
4.3 Installing in a vacuum system
• Connectionlinesatthepumpinlethavetobegastight.Particlesanddustmust
notbeaspirated,theuserhastoprovideappropriateltersifnecessary.Theuser
mustensuretheirsuitabilityconcerninggasow,chemicalresistanceandsafe-
ness against clogging prior to use.
• Connectanexhaustlinegastightatthepumpoutletifnecessary.Alwaysdispose
ofexhaustgasesappropriately(e.g.intoafumehood).Ifthereisriskofreleaseof
dangerousorpollutinguids,installanappropriatesystemtocatchanddispose
ofthoseuids.
• Reducethetransmissionofvibrationandpreventloadingduetorigidpipelines.
Insertelastichosesorexibleelementsascouplingsbetweenthepumpandrigid
pipes. Attention:Flexibleelementstendtoshrinkwhenevacuated.
• Assembletheglasssetstress-free.
• Theexhaustlinehasalwaystobefree(pressureless)toensureanunimpeded
discharge of gas.
• Especiallyifthegasballastvalveisopen,apowerfailuremaycauseuninten-
tional ventilation of the pump and the vacuum system. In case this constitutes a
potential source of danger, take appropriate safety measures.
• Makesureventilationisadequateifthepumpingunitisinstalledinahousingor
ifambienttemperatureiselevated.Provideexternalventingifnecessary.Keepa
distance of minimum 20cm between fan and ambient parts.
• Beforeswitchingoncheckthatmainsvoltageandcurrentconformwiththeequip-
ment (see rating plate).
If using the ROTAVAC vario control or Rotavac vario
tectogetherwiththeROTACOOLreuxcooler,place
the ROTAVAC vario control or ROTAVAC vario tec di-
rectly on top of the ROTACOOL.
If using the ROTAVAC vario control or ROTAVAC vario
tecwithouttheROTACOOLreuxcooler,anextrafoot
has to be installed for operation with the condensate
cooler (13; included in the scope of delivery of the con-
densate cooler). Position the ROTAVAC vario control
or the ROTAVAC vario tec simply on the extra foot,
thereby placing the pump’s feet in the four hollows of
theextrafoot.
Install the coolant hoses in such a way that the stability
of the pump is assured even with a full catchpot