Manual

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully
the validity of this document with respect to his product. Manual-no.: 01-005-005-36-1
page 30 of 35Translation of the original instructions
6.5 Replacing the overpressure safety relief device at the condenser
Overpressure safety relief device
...................................................................... 23-30-01-04-98
Round bottom ask 500 ml, coated
...................................................................... ...514-83000-02
Remove joint clip at the catchpot.
UnscrewthefourTorxscrewsatthecounterholderofthe
condenser and remove condenser. Thereby remove the
adapter from the inlet of the condenser.
Pull off old overpressure safety relief device and install
new one. Check for correct position of the PTFE-foil under
the overpressure safety relief device.
Reassembly in reverse order.
Position adapter in the inlet of the condenser and screw
condenser with counter holder to the pump (Torx screws).
Fix with union nut.
Reassemble catchpot.