page 16 of 35Translation of the original instructions
4 Use and operation
4.1 Installing in a vacuum system
➨ Ifthereisriskofreleaseofdangerousorpollutinguids,installanappropriate
systemtocatchanddisposeofthoseuids.
+ Connect an exhaust line gas tight at the pump outlet if necessary. Always dispose
of exhaust gases appropriately (e.g. into a fume hood).
+ Never block the gas outlet. The exhaust line must always be free of obstructions
(no back pressure) to ensure an unimpeded discharge of gas.
+ Particles and dust must not be aspirated, the user has to provide appropriate
lters if necessary.The user must ensure their suitabilityconcerning gas ow,
chemical resistance and safeness against clogging prior to use.
+ Make sure ventilation is adequate if the pumping unit is installed in a housing or
if ambient temperature is elevated. Provide external venting if necessary.
• Reducethetransmissionofvibrationandpreventloadingduetorigidpipelines.
Insertelastichosesorexibleelementsascouplingsbetweenthepumpandrigid
pipes. Attention: Flexible elements tend to shrink when evacuated.
• Connectionlinesatthepumpinlethavetobegastight.
• Especiallyifthegasballastvalveisopen,apowerfailuremaycauseuninten-
tional ventilation of the pump and the vacuum system. In case this constitutes a
potential source of danger, take appropriate safety measures.
• Beforeswitchingoncheckthatmainsvoltageandcurrentconformwiththeequip-
ment (see rating plate).
• Checkthat thevoltage selection switch is set correctly. Unplug the equipment
before setting the voltage selection switch! Note: If the equipment is switched on
with wrong voltage selection, the equipment may be damaged!
Keep a distance of minimum 20cm between fan and ambient parts.
Avoid throttling losses by using connecting pipes with large diameter and keeping
them as short as possible.
Install outlet pipelines always falling to avoid backow of condensate towards the
pump.
Useofasuitablevalvetoisolatethepumpfromthevacuumsystemisrecommended
to allow the pump to warm up before pumping condensable vapors or to clean the
pump before it is switched off.
When assembling, ensure vacuum-tightness. After assembly, check the whole sys-
tem for leaks.
Secure hose connections at the pump appropriately against accidental detaching.
If exhaust noise is annoying, connect an exhaust hose.
NOTICE
voltage selection switch
”115”: 90-126 V~
”230”: 180-253 V~