User Manual
Attachments Attachments
96 97
Technical data
Distimatic in general
Permissible ambient
conditions
5 - 31°C at 80 % relative humidity
32 - 40 °C linear decrease up to 50% relative humidity
0 - 2000 m a.s.l.
Contamination level 2
Overvoltage category II
Evaporation rates (L/h)
ΔT* 40°C
for Hei-VAP Distimatic
(continuous operation)
Water 1.2
Ethanol 3.5
Acetone 4.4
Toluene 6.3
Airborne sound level Signicantly lower than 80 dB(A)
Materials:
Collector Borosilicate glass
Ground joint parts PTFE
Laboratory screw joints PPS/PTFE
Collector valve/seal PTFE/FFKM
Evacuation valve/seal PVDF/FKM
Ventilation valve/seal PEEK/FFKM
Rell valve/seal PTFE/FFKM
Tube screw connections PVDF/FKM/PTFE
PTFE tubes PTFE; Ø 8x6 mm und Ø 6x4 mm
Check valve/seal PFA/FFKM
Filter PTFE; POR 1/G 1, nominal width of pores 100 - 160 µm
Sensor head PTFE
*
ΔT = Difference between heating bath temperature and boiling temperature
Control box
Voltage 230 V
Frequency 50/60 Hz
Measurements without
mounted components
(W x D x H)
300 x 200 x 120 mm (approx.)
Weight 5 kg (approx.)
Protection class IP54
Power input 50 VA
Connection Multiple switched socket (Hei-VAP) or multiple socket
(Hei-VAP Industrial)
Vacuum pump
Overow sensor for condensate and residue vessel
(only on Distimatic including automatic drainage of
residue)
3/2 way valve
Residue pump (only on Distimatic including automatic
drainage of residue)
Valve matrix
Cut-out sensor
Flask sensor
Operating panel
Wall bracket
Dimensions (W x D x H) 390 x 245 x 170 mm (approx.)
Weight 1 kg (approx.)