User Manual

Start-Up Start-Up
80 81
Operating states
Operating state Description
process stopped
Automatic mode is disabled.
Basic condition after activation of device or end of
automatic mode.
Status display Automatic mode is activated.
endprocess
Automatic mode in nal process.
If the rell sensor does not detect any medium during
the rell, the nal process is initiated. The remaining
media is distilled.
Following, the mains sockets for the rotary evaporator
and the vacuum pump are deactivated at the control
box. The vacuum circuit is ventilated. The condensate
pump pumps out the condensate.
Parameter overview
Set parameters – Sensor-Mode
All parameters of the following table are settable. The settable criteria of a parameter
(parameter settings) are marked in the following with the symbols „>…<”.
Select operational mode
After switching on the unit you can choose between two different operational modes.
Select Sensor-Mode by pressing the set key and the arrow key <.
Select Time-Mode by pressing the set key and the arrow key >.
Once you selected the Sensor-Mode a S will be visible on the display.
Once you selected the Time-Mode a T will be visilble on the display.
Changing modes is only possible after resetting the unit.
Parameter Description
Start/stop To gain access to the menu.
Rell to start
initial lling
a. The key rell starts the pre-programmed operation.
b. The key parameter allows for pre-setting all
parameters.
Time to rell >…< Time between relling of the medium. If the preset time
has expired, the ask sensor checks whether there is
still any medium in the evaporating ask and triggers
product relling if needed.
Extra time until drain off The residue in the evaporating ask is drained by the
residue pump if the evaporation process has stagnated
for the time period set in the parameter Extra time until
drain off.
Drain time
cond >…<
resi >…<
Cond = Drainage of condensate
Please enter the required time to drain condensate from
the collector
Resi = Drainage of residue
Please enter the required time to drain residue from the
evaporating ask.
ENDPROCESS
evaporate time >…<
Pre-program the time frame for the unit to power off all
periphery after the last cycle has been executed.
ENDPROCESS
draining
cond >…<
resi >…<
Prior to powering off, the collector (condensate) and the
evaporating ask (residue) can be drained. You have the
option to allow for a nal drainage (ON) or not (OFF).
Residue pump during
process >…<
To drain the residue this option must be set ON.
Rell to start
initial lling
Start the process by pressing the rell key.
Overow Collection
Vessel >…<
Activate the overow power-off by pressing the ON key
or not (OFF).
Mo 09:00
2006-01-27
Date and time display format. Here you can set the
current date and time. Since error codes are always
stamped with time, we recommend you to keep the date
up to date in order to facilitate error analysis.
Drainage of condensate or residue can be performed manually by pressing the rell /
drain key (condensate) and drain residue key (residue).