User Manual
Installation Installation
68 69
Optimization of the distillation process
Rell evaporating ask only up to one third. Position the ask sensor
accordingly.
Use large evaporating asks whenever possible.
Use the following recommended immersion depths:
The end of the PTFE tube (15) Ø 6 mm varies. PTFE tube (16) Ø 8
mm (only on Distimatic including automatic drainage of residue)
ends approximately 2 - 3 mm above the bottom of the evaporating
ask. Flask sensor (17) ends just above the required maximum ll
level.
This way your distillation process will become much faster.
The immersion depth of the PTFE tubes and the ask sensor in the evaporating ask
may be adjusted by loosening the respective screw cap slightly at the multiple screw
connection and by pulling or pushing the PTFE tube or the ask sensor.
Re-secure the screw caps on the multiple screw connection so they are hand-tight
again.
17
15
16
Connect cut-out sensor with rubber ring Ø 80 mm to distributor (18) (for glassware
set R) or condenser opening (19) (for glassware set A).
Cut-out sensor
18
19