User Manual
Installation Installation
54 55
Connect PTFE tube Ø 6 mm to
connection (1) and tighten until hand-
tight.
Route PTFE tubes and ask sensor into the evaporating ask and secure the multiple
screw connection at connection (4) until hand-tight.
Multiple screw connection
Route PTFE tube Ø 8 mm through connection (2) (only on Distimatic including
automatic drainage of residue) and PTFE tube Ø 6 mm through connection (3) and
secure in place so it is hand-tight.
The immersion depth of the PTFE tubes and the ask sensor in the evaporating ask
may be determined by loosening the respective screw cap slightly at the multiple
screw connection and by pulling or pushing the PTFE tube or the ask sensor.
Re-secure the screw caps on the multiple screw connection so they are hand-tight
again afterwards.
1
2
5
3
7
6
Optimization of the distillation process
Rell evaporating ask only up to one third. Position the ask sensor
accordingly.
Use large evaporating asks (3-5 liter variants) whenever possible.
Use the following recommended immersion depths:
The end of the PTFE tube (5) Ø 6 mm varies. Flask sensor (6) ends just
above the required maximum ll level. PTFE tube (7) Ø 8 mm (only on
Distimatic including automatic drainage of residue) ends approximately
2 – 3 mm above the bottom of the evaporating ask.
This way your distillation process will become much faster.
4