User Manual
Installation Installation
46 47
Condensate pump
Remove the screw caps (10) from connections (11) and (12). Connect PTFE tubes Ø 6 mm to connections (16) and (17) and tighten until hand-
tight.
Important: Observe ow direction!
Cut off the end (18) of the PTFE tube Ø 6 mm with a cutting knife only at an angle of
approx. 45° and route through the connections (20) and (19) of the overow sensor.
Suspend overow sensor safely on the edge
(21) of the condensate vessel.
The immersion depth of the PTFE tubes in
the vessel may be determined by pulling or
pushing the PTFE tube (22).
Ensure that the sheathed sensor wire is
not detached from the insulation during
installation of the PTFE tube.
Connect PTFE tubes Ø 6 mm to connections (13) and (14) and tighten until hand-
tight.
To integrate the check valve, vertically cut through the PTFE tube with the HOLEX
hose cutter at about point (15) shown in the circle.
Check valve in the condensate circuit
Overow sensor for condensate vessel
11
12
16
17
20
21
22
19
18
10
13
14
15
Condensate
Condensate
Condensate