Instructions / Assembly

27
Electromagnetic Compatibility
Table 2
Guidance and declaration of manufacturer-electromagnetic immunity
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electromagnetic environment
-guidance
Compliance
level
IEC 60601
test level
I
MMUNITY test
± 6 kV
contact
± 8 kV air
± 6 kV
contact
± 8 kV air
Electrostatic
discharge (ESD)
IEC 61000-4-2
Mains power quality should be
that of a typical commercial or
hospital environment.
±2 kV for
power supply
lines
±2 kV for
power suppl
y
lines
Electrical fast
transient/burst
IEC 61000-4-4
Mains power quality should be
that of a typical commercial or
hospital environment.
±1 kV line(s)
to line(s)
Surge
IEC 61000-4-5
<5 % U
T
(>95 % dip
in U
T)
for 0.5 cycle
40 % U
T
(60 % dip
in U
T)
for 5 cycles
70 % U
T
(30 % dip
in U
T)
for 25 cycles
<5 % U
T
(>95 % dip
in U
T)
for 5 s
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the device requires
continued operation during
power mains interruptions, it is
recommended that the device
be powered from an
uninterruptible power supply
or a battery.
<5 % U
T
(>95 % dip
in U
T)
for 0.5 cycle
40 % U
T
(60 % dip
in U
T)
for 5 cycles
70 % U
T
(30 % dip
in U
T)
for 25 cycles
<5 % U
T
(>95 % dip
in U
T)
for 5 s
Voltage dips,
shor
t
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
±1 kV line(s)
to line(s)
ELECTROMAGNETIC COMPATIBILITY