User's Manual
39
Note
*Note
!
This equipment has been tested and found to comply with the limits for a Class Bdigital device, pursuant to Part 15
of the FCC Rules. These limits are designed toprovide reasonable protection against harmful interference in a
residential installation.
This equipment generates uses and can radiate radio frequency energy and, if notinstalled and usedin accordance
with the instructions, may cause harmful interferenceto radio communications. However, there is no guarantee
thatinterference will not occur in a particular installation. If this equipment does causeharmful interference to radio
or television reception, which can be determined byturning the equipment off and on, the user is encouraged to try
to correct theinterference by one or more of the following measures:
The user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and the receiver.
Connect the equipment into an outlet on a circuit different from that to which thereceiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
CAUTION:
To assure continued FCC compliance:
1. Any changes or modifications not expressly approved by the grantee of this devicecould void the user's
authority to operate the equipment.
2. This equipment complies with FCC radiation exposure limits set forth for anuncontrolled environment. This
equipment should be installed and operated withminimum distance 20cm between the radiator & your body.
FCC Label Compliance Statement:
This device complies with Part 15 of the FCC Rules. Operation is subject to thefollowing two conditions:
(1) this device may not cause harmful interference, and
(2)this device must accept any interference received, including interference that may cause undesired operation.
*Note
!
“Changes or modifications not expressly approved by the manufacturer could void the user’s authority to operate
the equipment”.
HL858CC essential performance per IEC 80601-2-30 additional essential
performance requirements:
201.12.1.102 Limits of the error of the manometer from
environmental conditions
Over the temperature range of 5 °C to 40 °C (41 °F ~ 104 °F ) and
the relative humidityrange of 15 % to 93 %(non-condensing), the
maximum error for the measurement of theCUFF pressure at any
point of the NOMINAL measurement range shall be less than or
equalto ± 3 mmHg (± 0.4 kPa) or 2 % of the reading, whichever is
greater.
201.12.1.107 Limits of the change in error of the blood pressure
determination
The laboratory limits of the change in error of the BLOOD PRESSURE
DETERMINATION of theAUTOMATED SPHYGMOMANOMETER shall be
less than 3 mmHg (0.4 kPa).