Users Manual
24
HBM: Business Document
8 Functionality testing
You can check the
functionality
of the rotor and the
stator
from the LEDs on
the evaluation unit.
Fig. 8.1: LEDs on the evaluation unit
LED
Color
Significance
LED no. 2
(power status)
Green (permanently lit)
normal operation
The main switch (no. 6 in Fig. 6.1) is
on and the evaluation unit is under
power
Off (not lit)
The main power is off
LED no. 3
(RF level status)
Green (permanently lit)
normal operation
The LED indicates an acceptable RF
signal strength
Off (not lit)
The RF signal strength is too low, the
transmission between rotor and stator
is broken. Check rotor / stator
alignment
LED
(cal-function
status)
Off (not lit)
normal operation
The shunt signal is not active.
Orange (permanently
lit)
The shunt signal is active as long as
the LED is lit.
LED no. 2 (green),
power status
LED no. 3 (green),
RF level status
LED (orange),
cal-function status