User Manual

Hardy Floor Scales
22
Step 3. Use a small slotted head screw driver and
disconnect the Load Sensor cable. It is a
good idea to mark the wires if the wire
markers have been removed. Color code
wires are as follows:
We recommend the use of C2 cable from the
weighing instrument to the junction box. Note the
color codes for C2 load points (left to right facing the
rear panel). Match the load sensors color coding to
ease trouble shooting.
Load Sensor # Connector
Load Sensor #1 TB1
Load Sensor #2 TB2
Load Sensor #3 TB3
Load Sensor #4 TB4
To Instrument TB5
Table 3: Summing Card Load Sensor Connectors
Load cell connector
Rear panel 9 pin
C2 cable TB9 7 pin
to Junction Box
Load cell color codes
3 pin connector
Shield Ground Wire
C2- Violet
C2+ Grey
EXC- Black
SEN- Brown
SIG- White
SIG+ Green
SEN + BLUE
EXC+ RED
EXC- Black
SEN- Brown
EXC+ RED
SEN+ BLUE
Shield Ground Wire
C2+ Grey
C2- Violet
SIG+ Green
SIG- White
EXC+ RED
EXC- Black
Shield Ground Wire
4 pin connector
C2+ Grey
C2- Violet
SIG+ Green
SIG- White