User guide

Section 7: Operations
12/09 23 0596-0315-01 Rev B
Section 7 Operations
7.1. BASIC SCALE OPERATIONS
Upon power up the Indicator performs a warm-up cycle.
The Indicator initiates a test, displaying numbers 1 to 8, and lights up all LED’s.
The Program number and Revision Information displays.
The Indicator then displays the current weight on the scale
7.2. KEYPAD FUNCTIONS, WEIGH MODE
KEY ACTION
Recipe 1-4
Enables or disables programmed recipe in Checkweigh mode.
In peak weigh mode, 1 enables peak, 2 disables, and 3 resets peak.
TARE
Automatically tares off the displayed weight.
If held for 3 seconds, enables or disables accumulate function.
ZERO
Resets gross weight to center of zero.
PROGRAM/ENTER
Captures weight and adds to accumulation if enabled.
OVER/UNDER/NEXT
Displays accumulated weight and number of weighments if enabled.
If held for 3 seconds, enters Checkweigh or peak weight setup.
UNITS
Switches between selected weighting units.
B/G – NET
Toggles between gross and net weights.
If held for 10 seconds, displays audit trail functions.
PRINT
Transmits transaction data out com port 1 if button is selected.
7.3. INSTRUMENT WEIGHING FUNCTIONS
Basic Weighing
Ensure that the bench scale is empty, power on the indicator, press the zero key and the display indicates “0”
and is ready for use.
Gross Weighing
1. Press t
he b/g / net key, if required, to set the display to GR (gross).
2. Press the zero key, if required, to set scale to “0.”
3. Place container/object on the scale platform.
4. Read the gross weight on the display.
Net Weighing
1. Press t
he b/g / net key, if required, to set display to GR (gross).
2. Press the zero key, if required, to set scale to “0.”