User guide
Section 6: Serial I/O
12/09 12 0596-0315-01 Rev B
Cust: Custom units. The display will indicate cust and the current selection, either y for enable or n for
disable. Press over/under/next, units or tare to change the selection. Press Program/Enter to enter the
selection.
Note: When custom units are active, units LED indicators are off in the weigh mode.
Azt: Programming the Automatic Zero Tracking band: (this feature will maintain zero when small amounts of
material are placed on the scale, such as rain, snow, debris, etc.) The display will indicate azt x where x is the
current setting. Press over/under/next, units or tare to change the selection. Press Program/Enter to enter
the selection. Available selections are:
OFF Function is not active.
0.5 half of a division / increment / graduation
1 one division / increment / graduation
3
three divisions / increments / graduations
Bal: Programming the motion band (the range in divisions/increments/graduations that weight must be stable
before a print, zero, or tare function will be allowed). The display will indicate bal x where x is the current
setting. Press over/under/next, units or tare to change the selection. Press Program/Enter to enter the
selection. Available selections are:
OFF Function is not active.
0.5 half of a division / increment / graduation
1 one division / increment / graduation
3
three divisions / increments / graduations
O.r: Programming the zero range: (the percentage of scale capacity that may be removed by pressing the Zero
Key). The display will indicate o.r x where x is the current setting. Press over/under/next, units or tare to
change the selection. Press Program/Enter to enter the selection. Available selections are:
100
100 percent zero range
2 2 percent zero range
D: Programming the division size: The display will indicate d followed by the current division size setting.
Press Over/Under/Next or units to scroll up, or tare to scroll down through the available choices. Press
Program/Enter to enter the selection. Available selections are:
50 20 10 5 2 1
0.5 0.2 0.1 0.05 0.02 0.01
0.005 0.002 0.001 0.0005 0.0002 0.0001
Filter: Programming the filter setting: (intended to minimize the effects of motion, vibration, and wind
currents) The display will indicate filter followed by the current filter setting. Press Over/Under/Next or
units to scroll up, or tare to scroll down through the available choices. Press Program/Enter to enter the
selection. Available selections are:
1 3 5 11
15 20 30 50
Flf: Flush filter factor: Allows indicator to switch to fast filter rate if weight change exceeds the number of
divisions in setting. The display will indicate flf followed by the current setting. Press Over/Under/Next or
units to scroll up, or tare to scroll down through the available choices. Press Program/Enter to enter the
selection. Available selections are:
oFF
1 2 5 10 20 50 100
Tare: Programming the tare setting: (the means by which a container’s weight may be removed, to set the
indicator to display the net weight only) The display will indicate tare followed by the current tare setting.