O.r – zero tolerance range (set as % of capacity)
o 100
o 2
D – division size
o 50, 20, 10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05, 0.02, 0.01, 0.005, 0.002, 0.001,
0.0005, 0.0002, 0.0001
Filter – (intended to minimize effects of motion, vibration, and wind currents)
Flf – Flush filter. Allows switch to fast filter rate if weight changes quickly
Tare – enable tare function
o Off – tare inactive
o On – tare is active
o On-Clr – automatically clears when gross wt returns to zero.
Hl – overload limit (set as % of capacity)
o 102.5
o 103
o 105
o 110
o 300
Ul – underload limit (set as % of capacity below zero)
o 5.5
o 10
o 25
o 50
o 100
D rate – display update rate (set in tenths of seconds)
o From 0.1 to 1.0
P1 – serial port 1 output format
o GTN – if in gross mode will print gross weight. If in net mode, will print
gross/tare/net weights
o nEt – will print net only
P2 – serial port 2 output format
o GTN
o nEt
APP menu
X.xxupd – display of micro volts per division. Should be greater than 1.0. Read only.
If less than 1.0, adjust division size.
Acc – Enable/disable accumulator
A thld – Autoprint and accumulator threshold. Number of divisions to meet or exceed to
trigger auto print or accumulator.