User Manual
Table Of Contents
- Copyright Declaration
 - General Safety Summary
 - Safety Terms and Symbols
 - Product Scrapping
 - Contents
 - DSO2000 series digital storage oscilloscope introduction
 - 1. Quick start
 - 2. Function Introduction
- 2.1. Menu and Control Keys
 - 2.2. Connector
 - 2.3. Multi-function Knobs and Softkeys
 - 2.4. Oscilloscope Setup
 - 2.5. Horizontal Controls
 - 2.6. Vertical System
 - 2.7. Trigger System
 - 2.8. Protocol Decode
 - 2.9. Save/Recall
 - 2.10. Measure System
 - 2.11. DVM
 - 2.12. Acquire
 - 2.13. Display
 - 2.14. Utility System
 - 2.15. Fast Action Buttons
 
 - 3. Waveform Generator
 - 4. Remote Control
 - 5. Troubleshooting
 - 6. Services and Support
 - 7. General Care and Cleaning
 - Appendix A: Technical Specifications
 - Appendix B: Accessories
 - Appendix C Harmful and Poisonous Substances or Elements
 
  User Manual 
  30  /  72 
> (greater than a time value): trigger when the positive or negative pulse time of the input signal is 
greater than the specified time value.   
!= (not equal to a time value): trigger when the positive or negative pulse time of the input signal is not 
equal to the specified limit of time.   
= (equal to a time value): trigger when the positive or negative pulse time of the input signal is equal to 
the specified limit of time.   
6. Press the Time softkey and turn V0 to set the reference value of the time. 
7. Press the Mode softkey, turn V0 to select the trigger mode (auto, normal), and press V0 to confirm. 
Auto: When the oscilloscope meets the trigger condition, it completes a trigger acquisition once; when 
the trigger condition is not met, it can run the acquisition waveform freely. 
Normal: When the oscilloscope meets the trigger condition, the input waveform is displayed; when the 
trigger condition is not met, the original waveform is displayed. 
8. Press the Holdoff softkey and turn V0 to set the time that the oscilloscope waits before a trigger to 
the next trigger, so that complex waveforms are displayed stably. 
2.7.9. Under Amp Trigger   
The  Under  Amp  trigger  looks  for  pulses  that  cross  one  threshold  but  not  another  as  shown  in  the 
picture below. 
⚫   A positive under Amp pulse across through a lower threshold but not an upper threshold.   
⚫   A negative under Amp pulse across through an upper threshold but not a lower threshold.   
To trigger on under Amp pulse:   
1. Press the [Trig Menu] button on the front panel to enter the Trigger system function menu.   
2. Press the Type softkey, then turn the Multifunctional Knob to select Under Amp and push down the 
knob to confirm.   
3. Press the Source softkey, turn the Multifunctional Knob to select CH1~CH2 as the trigger source.   
4. Press the Polarity softkey to select Positive or Negative pulse to trigger.   
5. Press the When softkey, turn the Multifunctional Knob to select the desired condition (<, >, != or =).   
6. Press the Width softkey, and then turn the Multifunctional Knob to select the desired value.   