User Manual
Table Of Contents
- Copyright Declaration
 - General Safety Summary
 - Safety Terms and Symbols
 - Product Scrapping
 - Contents
 - DSO2000 series digital storage oscilloscope introduction
 - 1. Quick start
 - 2. Function Introduction
- 2.1. Menu and Control Keys
 - 2.2. Connector
 - 2.3. Multi-function Knobs and Softkeys
 - 2.4. Oscilloscope Setup
 - 2.5. Horizontal Controls
 - 2.6. Vertical System
 - 2.7. Trigger System
 - 2.8. Protocol Decode
 - 2.9. Save/Recall
 - 2.10. Measure System
 - 2.11. DVM
 - 2.12. Acquire
 - 2.13. Display
 - 2.14. Utility System
 - 2.15. Fast Action Buttons
 
 - 3. Waveform Generator
 - 4. Remote Control
 - 5. Troubleshooting
 - 6. Services and Support
 - 7. General Care and Cleaning
 - Appendix A: Technical Specifications
 - Appendix B: Accessories
 - Appendix C Harmful and Poisonous Substances or Elements
 
  User Manual 
  24  /  72 
= (equal to time value): trigger when the positive or negative slope time of the input signal is equal to 
the specified time value. 
For example, for a positive pulse, if you set t (pulse real width) = 200ns, the waveform will trigger. 
7. Press of 50% softkey to set the trigger level to the vertical midpoint between the peaks of the trigger 
signal. The trigger level values are displayed at the upper right corner of the screen. 
8. Press the Width soft key and select V0 to set the reference width of the pulse signal. 
The pulse width of the data source must be  ≥  5ns so that the oscilloscope can detect the pulse. 
=, ≠: Within a ±5% tolerance, triggers the oscilloscope when the signal pulse width is equal to or not 
equal to the specified pulse width.   
<,  >: Triggers the oscilloscope when the source signal pulse width is less than or greater than the 
specified pulse width. 
9. Press the Mode softkey, turn V0 to select the trigger mode (auto, normal), and press V0 to confirm. 
Auto: When the oscilloscope meets the trigger condition, it completes a trigger acquisition once; when 
the trigger condition is not met, it can run the acquisition waveform freely. 
Normal: When the oscilloscope meets the trigger condition, the input waveform is displayed; when the 
trigger condition is not met, the original waveform is displayed. 
10. Press the Holdoff softkey and turn V0 to set the time that the oscilloscope waits before a trigger to 
the next trigger, so that complex waveforms are displayed stably. 
2.7.3. Video Trigger   
Video  triggering  can  be  used  to  capture  the  complicated  waveform  of  most  standard  analog  video 
signals. The trigger circuitry detects the vertical and horizontal interval of the waveform and produces 
triggers based on the video trigger settings you have selected. The series of scope support standard 
video  signal field or  line of  NTSC  (National  Television  Standards Committee), PAL  (Phase Alternating 
Line).   
1. Press the [Trig Menu] button on the front panel to enter the Trigger function menu.   
2. Press the Type softkey, then turn the Multifunctional Knob to select Video and push down the knob 
to confirm.   
3. Press the Source softkey turn the Multifunctional Knob to select CH1~CH2 as the trigger source.   
4. Press Polarity to select the trigger polarity (positive and negative). 