INSTALLATION
When an outdoor installation is required, it is necessary to install a
transmitter to obtain accurate readings at distances from 10 to 50 m
(33-165').
Since the introduction of AmpHel
®
these distances are no longer a
problem. You can now connect your meter directly to an AmpHel
®
electrode, saving the cost of a transmitter.
The standard cable length of the AmpHel
®
electrode is 5 m (16.5').
Additional lengths of regular cable up to 50 m (165'), can be
installed without special connectors. Coaxial cables are always pre-
ferred over normal cables due to their excellent insulation, even
though, Amphel
®
electrodes can operate with both.
AmpHel
®
electrodes have a micro-amplifier in the electrode cap to
boost the signal, drastically reducing susceptibility to noise and drift.
With all of the components sealed in the electrode body, moisture up
to 100% RH will not effect the signal.
pH METER
WITH AN AmpHel ELECTRODE
MEASUREMENTS CAN BE
TAKEN FROM DISTANCES UP TO
50 METERS (165 FEET)
WET ENVIRONMENT
RELATIVE HUMIDITY UP TO 100%
MAXIMUM 50 METERS
®
pH-ORP METER
pH CALIBRATION SOLUTIONS
HI 7004M pH 4.01 buffer solution, 230 mL
HI 7004L pH 4.01 buffer solution, 460 mL
HI 7006M pH 6.86 buffer solution, 230 mL
HI 7006L pH 6.86 buffer solution, 460 mL
HI 7007M pH 7.01 buffer solution, 230 mL
HI 7007L pH 7.01 buffer solution, 460 mL
HI 7009M pH 9.18 buffer solution, 230 mL
HI 7009L pH 9.18 buffer solution, 460 mL
HI 7010M pH 10.01 buffer solution, 230mL
HI 7010L pH 10.01 buffer solution, 460 mL
ORP SOLUTIONS
HI 7020M 200-275mV ORP solution, 230 mL
HI 7020L 200-275mV ORP solution, 460 mL
HI 7091M Pretreatment reducing sol., 230 mL
HI 7091L Pretreatment reducing sol., 460 mL
HI 7092M Pretreatment oxidizing sol., 230 mL
HI 7092L Pretreatment oxidizing sol., 460 mL
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
The probe can be compensated for normal contamination by a process
of re-calibration. However, it is recommended that the process conduc-
tivity probe be removed from the system regularly for maintenance.
Rinse the HI 7637
probe with tap water. If a more thorough
cleaning is desired, unscrew the PVC sleeve and clean the sensors with
a non-abrasive cloth and alcohol. When reinstalling the PVC sleeve
onto the probe, be sure that the holes are towards the top of the
probe (the end with the cable).
After cleaning, re-calibrate the instrument. However, if the instrument
will not re-calibrate, replace the probe.
Note: Always re-calibrate the meter after the replacement of the
probe.