User Guide
8
Recorder output
The recommended recorder hookup uses a shielded, twisted-
pair cable. The shield should be connected to (earth) ground
at the recorder end and left open at the instrument end.
The output is 0.00 to 2.00 V corresponding to 0.00 to 2.00
µA probe current as indicated on the front panel display.
Power
Power is provided from the mains through a HI 710005 or HI
710006 power adapter. Be sure the mains voltage matches
the input voltage specified on the power adapter. The power
adapter output connector plugs into a socket on the back
panel of the instrument.
Caution: The power adapter may be damaged if not operated at the
correct voltage.
Fuse
The instrument power conditioning circuitry is protected with
a 200 mA, 5 x 20mm tubular fuse located on the rear panel.
To replace the fuse simply twist off the fuse holder cap and
replace the fuse.
Unplug the meter before replacing the fuse.