pHep® Pocket-sized pH Meter Untitled-1 1 9/10/2003, 10:03 AM
OPERATION: • Do not be alarmed if white crystals appear around the cap. This is normal with pH electrodes and they dissolve if rinsed with water. pH • Remove the protective cap and turn the pHep® on. • Immerse it in the sample up to the maximum immersion level. • Stir gently and wait until the display stabilizes. • After use, rinse the electrode with water to minimize contamination. • Store the electrode with a few drops of storage (HI 70300) or pH7 (HI 7007) solution in the protective cap.
• Allow the reading to stabilize and using a small screwdriver turn the pH 7 Calibration Trimmer to read 7.0. Calibration is now complete. ALWAYS USE FRESH BUFFERS FOR CALIBRATION & NEVER REUSE THEM. BATTERY REPLACEMENT: When the pHep® cannot be switched on or the display fades, pull out the battery compartment and replace all three 1.5V batteries, paying attention to their polarity. Batteries should only be replaced in a non-hazardous area using the battery types specified in this instruction manual.
pHep® Pocket-sized pH Meter SPECIFICATIONS: RANGE 0.0 to 14.0 pH RESOLUTION 0.1 pH ACCURACY ±0.1 pH (@20°C/68°F) TYPICAL EMC DEV. ±0.2 pH ENVIRONMENT 0 to 50°C (32 to 122°F); RH 95% non-condensing BATTERIES 3 x 1.5V alkaline LIFE approx.700 hours of use DIMENSIONS 150x30x24 mm (5.9x1.2x0.9") WEIGHT 85 g (3 oz.) ACCESSORIES: HI 77700P HI 7007L HI 7007M HI 70300L Calibration kit pH 7.01 (10 x 20 mL sachets) Buffer solution pH 7.01 (460 mL) Buffer solution pH 7.