User Guide
54
The colon is a useful indicator for the correct position of
squares.
Note A maximum of two keys may be pressed simultaneously to be
properly recognized.
To exit the keyboard test procedure press LCD, CAL and SETUP
simultaneously.
EEPROM SELFTEST
The EEPROM selftest procedure involves verifying the stored
EEPROM checksum. If the checksum is correct the “Stored
data good” message will be shown for a few seconds before
exiting selftest procedure.
If not, the message “Stored data error - Press to reset
stored data or to ignore”.
If is pressed the EEPROM selftest procedure terminates
with no other action. Otherwise, EEPROM is reset with de-
fault values from ROM as when a device with a virgin
EEPROM is switched on.
During EEPROM reset a blinking mes-
sage “Set MEM” is shown on the LCD.
At the end of this operation all the pa-
rameters are reset to their default
values. Calibration data is also reset.
For this reason the "CAL" flag blinks
until the pH calibration is performed.