User Guide
49
Note In any moment, by pressing LCD or CAL DATA the meter will
return to the regular operating display.
• Press or to view the time of last calibration. The sec-
ondary display will show "HOU" to indicate hours.
• Press or again to view the offset in mV at the time of
last calibration. The secondary display will show "OFF" to
indicate offset.
• Press or again to view the first slope in mV at the time
of last calibration. The secondary display will show "SL1" to
indicate first slope.
• Press or again to view the second slope in mV at the
time of last calibration. The secondary display will show
"SL2" to indicate second slope.
• Press or again to view the first memorized buffer at
the time of last calibration. The secondary display will show
"BUF1" to indicate first buffer.