User Guide
47
• Press CFM to confirm the parameter that stops blinking on
the primary display.The secondary display shows the multi-
meter or HI931002 input value as the interval lower limit.
• Use or to make the
HI931002 or multimeter output
correspond with the value shown
on the secondary display (e.g. 4).
• Wait until the calibrator reading is stable (approx. 30 sec.).
• Press CFM to enter. The meter will switch to the second
calibration point. Repeat the above procedure.
• After the desired readings are obtained, press CFM and
the meter will skip back to normal operating mode.
Note When adjusting values using the or it is important to
allow for sufficient response time (up to 30 seconds)
The table below lists the values of output codes along with
the calibration values (which are the analog output mini-
mum and maximum) as indicated on the display.
The secondary display indicates the current calibration value,
while primary display indicates the current calibration type.
OUTPUT CALIBRATION CALIBRATION CALIBRATION
TYPE CODE POINT 1 POINT 2
0-1 mA 0 0 mA 1 mA
0-20 mA 1 0 mA 20 mA
4-20 mA 2 4 mA 20 mA
0-5 VDC 3 0 VDC 5 VDC
1-5 VDC 4 1 VDC 5 VDC
0-10 VDC 5 0 VDC 10 VDC