• Use a Checktemp or a calibrated thermometer with a reso-
lution of 0.1° as a reference thermometer.
• Immerse the temperature probe in
the beaker with ice and water as near
to the Checktemp as possible.
• Press and hold first CFM and then
CAL to enter the temperature cali-
bration mode.
• Execute the password procedure.
• With pH 502, the meter will ask for the calibration proce-
dure code number. Use or to select code 1 for the
temperature calibration and press CFM to enter.
• CAL will blink on the LCD until the
meter confirms a steady reading.
• When the reading has stabilized at
a point near the first calibration
point, CAL will stop blinking and an
intermittent CFM will prompt the user
to confirm the first calibration.
• If the reading stabilizes at a reading
significantly variant from the first
setpoint, an intermittent WRONG
will prompt the user to check the
beaker or baths.
• After pressing CFM the unit will switch to the second cali-
bration point.
• Select 25 or 50°C by pressing or .
0 °C
(32 °F)
°
C