User Guide

45
Use a Checktemp or a calibrated thermometer with a reso-
lution of 0.1° as a reference thermometer.
Immerse the temperature probe in
the beaker with ice and water as near
to the Checktemp as possible.
Press and hold first CFM and then
CAL to enter the temperature cali-
bration mode.
Execute the password procedure.
With pH 502, the meter will ask for the calibration proce-
dure code number. Use or to select code 1 for the
temperature calibration and press CFM to enter.
CAL will blink on the LCD until the
meter confirms a steady reading.
When the reading has stabilized at
a point near the first calibration
point, CAL will stop blinking and an
intermittent CFM will prompt the user
to confirm the first calibration.
If the reading stabilizes at a reading
significantly variant from the first
setpoint, an intermittent WRONG
will prompt the user to check the
beaker or baths.
After pressing CFM the unit will switch to the second cali-
bration point.
Select 25 or 50°C by pressing or .
0 °C
(32 °F)
°
C