When the reading has stabilized at
a point near the first calibration
point, CAL will stop blinking and an
intermittent CFM icon will prompt the
user to confirm the first calibration.
If the display stabilizes at a value
significantly different from the first
setpoint, an intermittent WRONG
icon will prompt the user to check
and adjust the simulator and start
again.
After pressing CFM the unit will switch to the second cali-
bration point at 350 mV.
With mV 602 it is possible to select 1900 mV by pressing
or . After that, proceed as described above.
Note A measure is considered stable when it varies little within a
sequence of acquisitions. The number of acquisitions is fixed
so that the waiting time for blinking “CFM” is about 20 sec-
onds.
Calibration procedure may be interrupted by
pressing CAL. If the calibration procedure is in-
terrupted this way, or if the controller is switched
off before the last step, no calibration data is
stored to EEPROM.
TEMPERATURE CALIBRATION
The pH/mV controller is factory calibrated for the mV and
temperature inputs. However, the user may also perform a
temperature calibration.
• Prepare a beaker containing ice
and water at 0°C/32°F and an-
other one with hot water at 25°C/
77°F or 50°C/122°F.
0 °C
(32 °F)
°
C
50 ºC
(122 ºF)
°
C