OFFSET AND SLOPE DIRECT SELECTION
Whenever the pH electrode offset and slope parameters are
known, it is possible to directly calibrate the meter entering
the electrode parameters.
• Press the “CAL DATA” and then “SETUP”. The LCD will
show the default offset of -5.0 mV.
• Using the , and enter the electrode offset param-
eter (the value must be between -100 and +100 mV).
• Confirm the value by pressing CFM.
If offset is invalid the “WRONG”
indicator will blink on the LCD.
• The LCD will then show the default
slope for Hanna electrodes: 57.5
mV/pH.
Note If CAL DATA or LCD are pressed before CFM, calibration is
aborted without changing the data of previous calibration.
• Using the , and enter the electrode slope param-
eter (the value must be between +40 and +80 mV/pH).
• Confirm the value by pressing CFM. If
slope is invalid the “WRONG” indi-
cator will blink on the LCD.
Note Press LCD or CAL DATA to exit cali-
bration. The slope will be set to the
default value (57.5 mV/pH).
Note Calibration data can be viewed pressing CAL DATA while in
normal mode but no pH buffer values will be displayed dur-
ing the data scrolling on LCD.