User Guide

42
OFFSET AND SLOPE DIRECT SELECTION
Whenever the pH electrode offset and slope parameters are
known, it is possible to directly calibrate the meter entering
the electrode parameters.
Press the “CAL DATA” and then “SETUP”. The LCD will
show the default offset of -5.0 mV.
Using the , and enter the electrode offset param-
eter (the value must be between -100 and +100 mV).
Confirm the value by pressing CFM.
If offset is invalid the “WRONG
indicator will blink on the LCD.
The LCD will then show the default
slope for Hanna electrodes: 57.5
mV/pH.
Note If CAL DATA or LCD are pressed before CFM, calibration is
aborted without changing the data of previous calibration.
Using the , and enter the electrode slope param-
eter (the value must be between +40 and +80 mV/pH).
Confirm the value by pressing CFM. If
slope is invalid the “WRONG” indi-
cator will blink on the LCD.
Note Press LCD or CAL DATA to exit cali-
bration. The slope will be set to the
default value (57.5 mV/pH).
Note Calibration data can be viewed pressing CAL DATA while in
normal mode but no pH buffer values will be displayed dur-
ing the data scrolling on LCD.