User Guide

40
CALIBRATION WITH MANUAL TEMPERATURE COMPENSATION
Enter the calibration procedure and press LCD to display
the temperature on the secondary LCD.
Unplug any temperature probe that may
be attached to the meter. The "°C" sym-
bol will flash.
• Note the temperature of the buffer solutions with a
ChecktempC or an accurate thermometer with a resolution
of 0.1°C.
•Use or to manually adjust the display reading to the
value of the reference thermometer (e.g. 20°C).
Follow the calibration procedure above.
Note To toggle between the pH buffer and
the temperature press LCD.
When a one-point calibration is carried out only the pH off-
set is computed and stored, while the pH slope is fixed
according to the theoretical values.
With a two-point calibration, offset and slope are computed
to fit the two calibration points. With a three-point calibra-
tion the offset and first slope values refers to pH 4.01 and
7.01 buffers, while the second slope refers to pH 7.01 and
10.01 buffers.
Note If the process meter has never been cali-
brated or an EEPROM reset has occurred,
the meter continues to perform measure-
ment. However, the user is informed of a pH calibration
requirement by a blinking “CAL” (see “Startup” section).
The device must be calibrated within the temperature range
of 0-95°C. Outside this range, the buffer pH values are not
reliable.