User Guide

39
Three-point Calibration
Proceed as described above but do
not quit calibration by pressing CAL.
Note The meter will automatically skip the two buffers that were
used to prevent errors.
After the second calibration point
is confirmed, immerse the pH elec-
trode and the Potential Matching Pin
into the third buffer solution (e.g.
pH 10.01) and stir gently.
Note The electrode should be submerged
approximately 4 cm (1½") in the so-
lution. The temperature probe should
be located as close as possible to the
pH electrode.
Only when the reading is stable the
probe indicator " " will stop flash-
ing (after about 30 seconds) and the
"CFM" indicator will start blinking.
Press CFM to confirm the calibration; if the reading is close
to the selected buffer, the meter stores the reading, adjust-
ing the 2
nd
slope point and the calibration process is ended
with the offset and the 1
st
and 2
nd
slope of the meter cali-
brated.
If the reading is not close to the se-
lected buffer, "WRONG
BUF
" will
blink.
Note During calibration, the secondary LCD displays the selected
buffer value. By pressing LCD the temperature can be dis-
played. This will allow you to check the buffer temperature
during calibration.