User Guide
39
Three-point Calibration
• Proceed as described above but do
not quit calibration by pressing CAL.
Note The meter will automatically skip the two buffers that were
used to prevent errors.
• After the second calibration point
is confirmed, immerse the pH elec-
trode and the Potential Matching Pin
into the third buffer solution (e.g.
pH 10.01) and stir gently.
Note The electrode should be submerged
approximately 4 cm (1½") in the so-
lution. The temperature probe should
be located as close as possible to the
pH electrode.
• Only when the reading is stable the
probe indicator " " will stop flash-
ing (after about 30 seconds) and the
"CFM" indicator will start blinking.
• Press CFM to confirm the calibration; if the reading is close
to the selected buffer, the meter stores the reading, adjust-
ing the 2
nd
slope point and the calibration process is ended
with the offset and the 1
st
and 2
nd
slope of the meter cali-
brated.
If the reading is not close to the se-
lected buffer, "WRONG
BUF
" will
blink.
Note During calibration, the secondary LCD displays the selected
buffer value. By pressing LCD the temperature can be dis-
played. This will allow you to check the buffer temperature
during calibration.