User Guide

17
CONTROL MODE
The control mode is the normal operational mode for these
meters. During control mode pH 502 and mV 602 fulfill the
following main tasks:
convert information from pH/ORP and temperature inputs
to digital values;
control relays and generate the analog outputs as deter-
mined by the setup configuration, display alarm condition;
RS485 management (if the feature is included).
In addition, pH 502 and mV 602 can log working data
through RS485 connection. This data includes:
pH, mV and
o
C measured values;
last calibration data;
setup configuration (also from PC).
The status of the meter is shown by the LED’s on the right
STATUS LEDs
Control Alarm Alarm LED (green) Relay LED (yellow) Red LED
OFF ---- ON OFF ON
ON OFF ON ON or OFF OFF
ON ON OFF ON or OFF Blinking
Meter exits control mode by pressing SETUP or CAL and con-
firming the password. Note that this command generates a
temporary exit. To deactivate the con-
trol mode definitively, set CONTROL
ENABLE to “0” (item # 02).
RELAY MODES
Once enabled, the relays 1 and 2 can be used in four differ-
ent modes):
1) ON/OFF, high setpoint (acid dosage) (not for pH5025YZ
and mV6025YZ);
2) ON/OFF, low setpoint (base dosage) (not for pH5025YZ
and mV6025YZ);
3) PID, low setpoint (base dosage, if available);
4) PID, high setpoint (acid dosage, if available).