User Guide
37
• Press CFM to confirm the selected parameter that will stop
blinking on the primary display. The secondary display shows
the HI 931002 or multimeter input value as lower limit of
the interval.
• Use the or to make the
HI 931002 or multimeter out-
put correspond with the
meter’s value shown on the
secondary display (e.g. 4).
• Wait for approximately 30 seconds (until the reading of
the calibrator is stable).
• Press CFM to enter. The meter will switch to the second
calibration point. Repeat the above procedure.
• After the desired readings are obtained, press CFM and
the meter will skip back to normal operating mode.
Note When adjusting values using the or it is important to
allow for sufficient response time (up to 30 seconds)
The table below lists the values of output codes along with
the calibration point values (which are the analog output
minimum and the analog output maximum) as indicated
on the display.
The secondary display indicates the current calibration point
value, while primary display indicates the current calibra-
tion type.