User Guide

26
Excel
©
Copyright of “Microsoft Co.”
Lotus 1-2-3
©
Copyright of “Lotus Co.”
Windows
®
and Windows Terminal
®
are registered Trademark of “Microsoft Co.”
Note Cables other than HI 920010 may use a different configu-
ration. In these cases, no communication between the meter
and the PC is possible. If you are not using the HI 920010
cable, contact the nearest Hanna Customer Service Center
or proceed as follows for a proper electrical connection:
pH 500/mV 600 PC
9-pin DSUB male connector 9-pin DSUB female connector
Pin 2 Pin 3 (Txd)
Pin 3 Pin 2 (Rxd)
Pin 4 Pin 6 (Txd)
Pin 5 Pin 5 (Gnd)
Pin 6 Pin 4 (DTR)
Pin 7 short circuit with 8
(RTS+CTS)
pH 500/mV 600 PC
9-pin DSUB male connector 25-pin DSUB female connector
Pin 2 Pin 2 (Txd)
Pin 3 Pin 3 (Rxd)
Pin 4 Pin 6 (Txd)
Pin 5 Pin 7 (Gnd)
Pin 6 Pin 20 (DTR)
Pin 4 short circuit with 5
(RTS+CTS)
SETTING THE BAUD RATE
The transmission speed (baud rate) of the meter and the
external device must be identical.
The meter is factory set to 4800. If you wish to change this
value, use item 71 in the setup mode (see page 14).