User Guide

17
CONTROL MODE
The control mode is the normal operational mode for these
meters. During control mode pH 500 and mV 600 fulfill the
following main tasks:
convert information from pH/ORP and temperature inputs
to digital values;
control relays and generate the analog outputs as deter-
mined by the setup configuration, display alarm condition;
RS232 management.
In addition, pH 500 and mV 600 can log working data
through RS232 connection. This data includes:
pH, mV and
o
C measured values;
last calibration data;
setup configuration (also from PC).
The status of the meter is shown by the LED’s on the right
STATUS LEDs
Control Alarm Alarm LED (green) Relay LED (yellow) Red LED
OFF ---- ON OFF ON
ON OFF ON ON or OFF OFF
ON ON OFF ON or OFF Blinking
Meter exits control mode by pressing SETUP or CAL and con-
firming the password. Note that
this command generates a tem-
porary exit. To deactivate the
control mode definitively, set
CONTROL ENABLE to “0” (item # 02).
RELAY MODES
Once enabled, the relays 1 and 2 can be used in four differ-
ent modes:
1) ON/OFF, high setpoint (acid dosage);
2) ON/OFF, low setpoint (base dosage);
3) proportional, low setpoint (base dosage, if available);
4) proportional, high setpoint (acid dosage, if available).