User Guide
16
then S1>S2, S2–D2>LA, HA>S1+D1;
If M1 = 4 and M2 = 3
then S2>S1, S1–D1>LA, HA>S2+D2;
where the minimum deviation (D1 or D2) is 0.5 pH (for
pH 500) or 25 mV (for mV 600).
Note The password cannot be viewed when SETUP is pressed
without entering the original password first. The default
password is set at “0000”. In the event that the user forgets
the password, this can be reset to “0000” by pressing and
holding CFM and then pressing LCD and CAL DATA at the
same time when the pH controller is in normal operating
mode (idle or control with measurement displaying).
Note When a wrong setup value is con-
firmed, the pH controller does not
skip to the next setup item but remains
in the current item displaying a flash-
ing “WRONG” indicator until the pa-
rameter value is changed by the user (the same is also true
for the setup code selection). In some circumstances, user
cannot succeed in setting a parameter to a desired value if
the related parameters are not changed beforehand; e.g. to
set a pH high setpoint to 10.00 the high alarm must be set to
a value greater than pH 10.00 first.
Note For code numbers 40, 41, 42, the output is related to pH or
mV units depending on the model (pH process meters or mV
process meters).