Code Valid Values Default PW
92 EEPROM selftest 0: off 0 yes
1: on
93 Relays and LEDs selftest 0: off 0 yes
1: on
94 Watchdog selftest 0: off 0 yes
1: on
99 Unlock password 0000 to 9999 0000 yes
Note The process controller automatically checks to ensure that
the entered data matches other related variables. If a wrong
configuration is entered, “WRONGblinks on the LCD to
prompt the user. The correct configurations are the following:
If M1=/ 0 then S1<HA, S1>LA;
If M2=/ 0 then S2<HA, S2>LA;
If M1= 1 then S1-H1>LA;
If M1= 2 then S1+H1<HA;
If M1= 3 then S1+D1<HA;
If M1= 4 then S1-D1>LA;
If M2= 1 then S2-H2>LA;
If M2= 2 then S2+H2<HA;
If M2= 3 then S2+D2<HA;
If M2= 4 then S2-D2>LA;
If M1= 1 and M2 = 2
then S1-H1>S2+H2, S2>LA, HA>S1;
If M1 = 2 and M2 = 1
then S2-H2>S1+H1, S1>LA, HA>S2;
If M1 = 3 and M2 = 2
then S1>S2+H2, S2>LA, HA>S1+D1;
If M1 = 2 and M2 = 3
then S1+H1<S2, S1>LA, HA>S2+D2;
If M1 = 4 and M2 = 1
then S1<S2–H2, S1–D1>LA, HA>S2;
If M1 = 1 and M2 = 4
then S1–H1>S2, S2–D2>LA, HA>S1;
If M1 = 3 and M2 = 4