Wait for about 30 seconds. Press CFM.
The secondary LCD section will show
"50.0°C".
Immerse the temperature
probe in the vessel with hot
water.
Wait for about 30 seconds.
Press CFM.
Complete your temperature
calibration by pressing CAL
if you do not want to per-
form the mV calibration (see
page 44).
PROCEDURE FOR pH 301
Switch the pH301 on while pressing CFM.
The "CAL" indicator will be lit. The sec-
ondary LCD section will show "0.0°C".
Immerse the temperature
probe in the vessel with the
ice and water and wait for a
few minutes.
CFM
°C
CAL
50 °C
(122 °F)
°
C
CAL
°C
CAL
CFM
0 °C
(32 °F)
°
C
CFM
ON
(for Technical Personnel only)
The pH meter has been factory calibrated for
the temperature and is ready for measure-
ments.
The temperature probes are interchangeable
and no temperature calibration is needed when
the temperature probe is replaced.
If, for any reason, the temperature measure-
ments are inaccurate, re-calibration should
be carried out as follows.
INITIAL PREPARATION
Prepare a
vessel con-
taining ice (at
0.0°C/32°F)
and water and
another one
containing hot
water (at a temperature of 50.0°C/122°F).
Place insulation material around the con-
tainer to minimize temperature changes.
Use a ChecktempC or a calibrated ther-
mometer with a resolution of 0.1 °C as a
reference thermometer.
PROCEDURE FOR pH 300
Switch the pH 300 on while pressing CAL.
The "CAL" indicator will be lit. The sec-
ondary LCD section will show "0.0°C".
Immerse the temperature
probe in the vessel with the
ice and water.
TEMPERATURE CALIBRATION
0 °C
(32 °F)
°C
50 ºC
(122 ºF)
°C
0 °C
(32 °F)
°
C
°C
CAL
ON
CAL