User Guide

4544
Only when the read-
ing is stable, the "
BUF
1
"
indicator will stop
flashing (after about
30 seconds) and the
"CFM" indicator will
blink. The meter will check for the slope
reading.
Press CFM to confirm the
calibration.
The meter stores the read-
ing (and adjusts the slope
point).
The ION calibration process is complete and
the meter will return to the normal operating
mode.
Note: the meter will automatically skip the buffer
that was used for the first calibration to
avoid erroneous calibration.
Note: during calibration, the sec-
ondary LCD displays the
selected calibration value.
By pressing RANGE the
temperature value can be
also displayed. This will al-
low you to check the solu-
tion temperature during cali-
bration.
CFM
CFM/
RANGE
Press CAL and the calibra-
tion process is ended with
only the offset of the meter
calibrated. For best accuracy
however, it is recommended
that a two-point calibration is
performed.
Note: during calibration, the sec-
ondary LCD displays the
selected buffer value. By
pressing RANGE the tem-
perature value can be also
displayed. This will allow you
to check the buffer tem-
perature during calibration.
TWO-POINT CALIBRATION
A two-point calibration is recommended for
best accuracy.
Proceed as described above for one point
calibration but do not quit calibration by
pressing CAL at the end.
After the first calibration point is confirmed,
immerse the ISE electrode into the sec-
ond calibration solution (e.g. 100 ppm)
and shake briefly.
Note: the electrode should
be submerged ap-
proximately 4 cm
(1½") into the solu-
tion.
Press UP or DOWN until the secondary
display changes the value of the second
solution (e.g. 100).
CAL
RANGE
4 cm
(1
½
")
°C
°C
2
ppm
CAL
BUF