Instruction Manual pH 209-209R Bench-top pH Meter http://www.hannainst.
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. This instrument is in compliance with directives. TABLE OF CONTENTS PRELIMINARY EXAMINATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 GENERAL DESCRIPTION . . . . . . . . .
FUNCTIONAL DESCRIPTION SPECIFICATIONS Range Resolution Accuracy (@20°C/68°F) Typical EMC Deviation Calibration pH mV pH mV pH mV pH mV Temperature Compensation Electrode (included) Analog output (PH09R only) Input Impedance Power 1) Liquid Crystal Display (LCD) 2) pH slope adjustment knob 3) pH offset adjustment knob 4) pH range selection key 5) mV range selection key 6) C range selection key 7) Temperature setting knob 8) ON/OFF switch 9) Power adapter socket 10) BNC electrode connector 11) Electrode
OPERATIONAL GUIDE • Press the pH key to display the pH measurement. Power connection Plug the 12VDC adapter (HI710005 or HI710006) into the power supply socket (#9 on page 4). Note: make sure the main line is protected by a fuse. Electrode connection For combination pH or ORP electrodes (with internal reference) connect the electrode's BNC to the socket provided (#10 on page 4).
p H CALIBRATION • Note the temperature of the buffer solution using a ChecktempC or a glass thermometer (e.g.20 C). IMPORTANT The instrument's pH range should be re-calibrated: • When the meter is new. • Whenever the pH electrode is replaced. • At least once a month. • After use in aggressive chemicals. • After cleaning procedure and changing the reference electrolyte. • For greatest accuracy. • Press the C key to select temperature setting.
PH VALUES AT VARIOUS TEMPERATURES Temperature has an effect on pH. The calibration buffer solutions are affected by temperature changes to a lesser degree than normal solutions. For manual temperature calibration please refer to the following chart. TEMP °C °F 32 0 41 5 10 50 15 59 20 68 25 77 30 86 35 95 40 104 45 113 50 122 55 131 60 140 65 149 70 158 4.01 4.01 4.00 4.00 4.00 4.00 4.01 4.02 4.03 4.04 4.05 4.06 4.07 4.09 4.11 4.12 6.86 6.98 6.95 6.92 6.90 6.88 6.86 6.85 6.84 6.84 6.83 6.83 6.84 6.84 6.
For refillable electrodes: If the filling solution (electrolyte) is more than 2½ cm (1") below the fill hole, add HI 7082 or HI 8082 3.5M KCl Electrolyte Solution for double junction or HI 7071 or HI 8071 3.5M KCl+AgCl Electrolyte Solution for single junction electrodes. For a faster response, unscrew the fill hole screw during measurements. For AmpHel®electrodes: If the electrode does not respond to pH changes, the battery is run down and the electrode should be replaced.
TEMPERATURE-RESISTANCE CORRELATION FOR HANNA pH SENSITIVE GLASS The resistance of glass electrodes partially depends on the temperature. The lower the temperature, the higher the resistance. It takes longer time for the reading to stabilize if the resistance is higher. In addition, the response time will suffer to a greater degree at temperatures below 10 C.
HI 1330B Glass-body, semimicro, single junction, refillable, combination pH electrode. Use: laboratory, vials. ORP PRETREATMENT SOLUTIONS HI 7091L Reducing Pretreatment Solution, 460 mL HI 7092L Oxidizing Pretreatment Solution, 460 mL pH ELECTRODES All electrodes part numbers ending in B are supplied with a BNC connector and 1 m (3.3') cable. 5mm DIA 0.2" 5mm 0.2" HI 1330 120 mm 4.7" "S" VERSION HI 1043B Glass-body, double junction, refillable, combination pH electrode. Use: strong acid/alkali. 9.
FC 100B Plastic-body (Kynar®), double junction, refillable, combination pH electrode. Use: general purpose for food industry. HI 1413B Glass-body, single junction, flat tip, Viscolene, non-refillable, combination pH electrode. Use: surface measurement. 12 mm 0.5" 12 mm 0.5" FC 100 HI 1413 120 mm 4.7" 110 mm 4.3" FC 200B Plastic-body (Kynar®), open junction, conic, Viscolene, non-refillable, combination pH electrode. Use: meat & cheese. 6 mm 0.
Extension cables for screw-type electrodes (screw to BNC adapter) HI7855/1 Extension cable 1m (3.3') long HI7855/3 Extension cable 3m (9.9') long HI7855/5 Extension cable 5m (16.5') long HI7855/10 Extension cable 10m (33') long HI7855/15 Extension cable 15m (49.5') long HI 7855 SERIES CABLE CONNECTORS CONNECTOR AND 3.0 mm (0.
WARRANTY CE DECLARATION OF CONFORMITY All Hanna Instruments meters are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrodes and the probes are guaranteed for a period of six months. This warranty is limited to repair or replacement free of charge. Damage due to accident, misuse, tampering or lack of prescribed maintenance are not covered.
HANNA LITERATURE Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The reference literature currently covers areas such as: • • • • • • • Water Treatment Process Swimming Pools Agriculture Food Laboratory Thermometry and many others. New reference material is constantly being added to the library. For these and other catalogs, handbooks and leaflets, contact your dealer or the Hanna Customer Service nearest to you.