Users Manual

51
PROBE MAINTENANCE
18. PROBE MAINTENANCE
Cleaning the HI764113 Probe Body
Rinse the probe with clean water to remove debris from around the probe body; wipe with a soft
cloth. Unscrew and remove the protective shield and set aside. Ensure weeds, debris, or other
materials are not coating the protective guard as this will block sample circulation preventing
sample from reaching the active measurement surface.
The exterior of the HI764113 can be cleaned by wiping the exterior surface with a aqueous-
soapy mixture and gently rubbing the exterior surface. Rinse with clean water.
The Stainless steel guard can be cleaned and polished with a cleaner suitable for stainless steel
cookware or appliances. Avoid chlorine bleach containing products.
Inspection
• Routinely Inspect probe for biofouling. Carefully inspect the probe body and Smart Cap.
A scratch in the black protective layer on the Smart Cap will affect the calibration (and
measurement). Replace the Smart Cap if the sensing surface has been compromised.
Cleaning the Smart Cap
• Use a mild detergent and a soft bristled toothbrush to clean the Smart Cap.
Rinse the cap with water after cleaning and dry with a laboratory tissue.
Yearly Replacement of the Smart Cap
Easy to use smart caps contain pre-loaded calibration coefficients that are automatically
transmitted to the probe. The Smart Cap stores data in a RFID tag. If caps are switched between
probes, no information will be lost. Smart Cap installation dates are easily tracked via the Probe
info screen and when powering , the meter displays remaining cap life.
Rinse the probe with clean water to remove debris from around the probe body; wipe with a
soft tissue.
opdo
®
Probe Replacement Cap Kit
HI764113-1
Contains:
1 opdo probe Smart Cap
6 g sachet with silicone grease
syringe
1 lens wipe
1 certification/instruction sheet
Note: Verify time and date are properly set on meter prior to new cap initialization.