SETUP
8. SETUP
Setup menu allows viewing and modifying the measurement parameters.
The following table lists the Setup parameters, their valid range and the factory default settings.
Item Description Valid value Default
Log Separator Log file column separator Comma, Semicolon Comma
Salinity The solution’s salt content 0 to 70 PSU 0 PSU
Manual Pressure Used to manually set the pressure
and disable the automatic
barometer measurement
Enabled or disabled Disabled
Pressure Unit mmHg, inHg, atm,
mbar, psi, kPa
mmHg
Temperature Unit °C or °F °C
Calibration Timeout Calibration due timer Disabled, 10 to
70 days
Disabled
BOD Configuration
Sample min delta DO The minimum diff. between the
start and the end DO val.
0.00 to 50.00 mg/L 0.00 mg/L
Sample min end DO The minimum end DO value 0.00 to 50.00 mg/L 0.00 mg/L
Seed min delta DO The minimum diff. between the
start and the end DO val.
0.00 to 50.00 mg/L 0.00 mg/L
Seed min end DO The minimum end DO value 0.00 to 50.00 mg/L 0.00 mg/L
OUR Configuration
Min time The minimum time for the OUR test 1 to 3600 s 1 s
Max time The maximum time for the OUR test 1 to 3600 s 3600 s
Min start DO The minimum DO value for starting
the OUR test
0.01 to 50.00 mg/L 0.01 mg/L
Min end DO The minimum DO value at the end
of the OUR test
0.00 to 50.00 mg/L 0.00 mg/L
Total volume The total volume of the solution to
be tested
0.1 to 300.0 mL 0.1 mL
Sample volume The volume of sample in the
solution to be tested
0.1 to 300.0 mL 0.1 mL