Calibration is a 1-point procedure. Selectable calibration points are:
0.0, 84.0µS, 1413µS, 5.00mS, 12.88mS, 80.0mS,111.8mS.
To enter EC calibration select the EC range and press
the CAL key.
Note: TDS reading is automatically derived from the EC reading and
no specific calibration for TDS is needed. Pressing CAL while
TDS range is selected has no effect.
Rinse the probe with some of the calibration solution or
deionized water. Immerse the probe into the solution.
The sleeve holes must be completely submerged. Tap
the probe repeatedly to remove any air bubbles that
may be trapped inside the sleeve.
For zero calibration, just leave the dry probe in air.
The indications "BUF" and "CAL" are dis-
played. The upper LCD shows the uncalibrated
EC reading. The lower LCD shows the buffer
value. The stability indicator "
" blinks.
Select the desired value with the V
and W keys, if necessary.
When the " " symbol stops blinking, the
reading is stable. The "CON" indication starts
blinking on the LCD asking for confirmation.
Press ALT + CFM to confirm the calibra-
tion.
If everything is satisfactory, the meter displays the "Stor Good" mes-
sage and returns to measurement mode.
Notes
I If the uncalibrated reading is too far from the expected value,
calibration is not recognized. The "CON" indication does not
appear; the " " and "BUF" symbols blink to signal wrong or
contaminated calibration solution.
II For best results choose an EC buffer value close to the sample
to be measured.