User Guide
8
TEMPERATURE COMPENSATIONTEMPERATURE COMPENSATION
TEMPERATURE COMPENSATIONTEMPERATURE COMPENSATION
TEMPERATURE COMPENSATION
Three options of compensating temperature are available:
Automatic (Atc): The probe has a built-in temperature sensor; the
value of the temperature is used to automatically compensate the EC/
TDS reading using 25C as reference temperature. This is the default
option.
Manual (Mtc): The temperature value, shown on the lower LCD, can
be manually set with the arrow keys. The compensation is referenced
at 25C. The "C" symbol blinks when this option is active.
No Compensation (notc): The temperature value shown on the lower
LCD is not taken into account. The reading displayed on the upper LCD
is the actual EC or TDS value. The "C" and "%TC" symbols blink when
this option is active.
To select the desired option press the ATC key until
the option is briefly displayed on the LCD.
Note: The default compensation mode is ATC.
If temperature compensation is active, measurements are compen-
sated using a default temperature coefficient of 1.90 %/C.
It is possible to select a different temperature coefficient (TC) in the
0.00 to 6.00 %/C range. To change the TC factor enter the setup
mode and select the "tc" item (see SETUP for details).
The current temperature coefficient can be
quickly viewed pressing ALT+TC. The value
is briefly displayed on the lower LCD.
The EC and TDS scales are autoranging. The meter automatically sets
the scale with the highest possible resolution.
By pressing ALT+FIXED, the autoranging
feature is disabled and the current range
is frozen on the LCD. "F1" symbol blinks
on the LCD.
To restore the autoranging option press ALT+FIXED again.
Note: Autoranging is automatically restored if the range is changed, if
the setup or calibration modes are entered and if the meter is
turned off and back on again.