User Guide
7
Press the ON/OFF key to turn the meter on.
Immerse the probe into the solution to be tested.
The sleeve holes must be completely submerged.
Tap the probe repeatedly to remove any air bubbles
that may be trapped inside the sleeve.
If needed, press the RANGE key repeatedly until the
desired range (EC, TDS, NaCl) is selected on the
LCD.
Allow for the reading to stabilize. The upper
LCD displays the measure in the selected
range while the temperature is displayed on
the lower LCD.
Notes
I If the meter displays "----" the reading is out of range.
II If the reading is unstable, the stability indicator "
" blinks.
III The "gm" indication on the LCD means g/L.
IV Make sure the meter is calibrated before taking measurements.
V If measurements are taken successively in different samples, to
have accurate readings it is recommended to rinse the probe
thoroughly with deionized water before immersion in the samples.
VI TDS reading is obtained multiplying the EC reading by the TDS
factor, which has a default value of 0.50. It is possible to
change the TDS factor in the 0.40 to 0.80 range by entering
the setup mode and selecting the "tdS" item (see SETUP for
details).
VII When the use of an alternate function (FNC, CFM,FIXED,TC and
CALT) is requested, press and hold the ALT key first and then
the second key.
TAKING MEASUREMENTSTAKING MEASUREMENTS
TAKING MEASUREMENTSTAKING MEASUREMENTS
TAKING MEASUREMENTS