PROBE MAINTENANCEPROBE MAINTENANCE
PROBE MAINTENANCEPROBE MAINTENANCE
PROBE MAINTENANCE
Rinse the probe with clean water after measurements. If a more
thorough cleaning is required, remove the probe sleeve and clean the
probe with a cloth or a nonabrasive detergent. Make sure to reinsert
the sleeve onto the probe properly and in the right direction. After
cleaning the probe, recalibrate the instrument.
The platinum rings support is made of glass. Take great care while
handling the probe.
Connect the meter
to a PC through the RS232 output (#11 on page
4). Use HI 920010 (9 to 9-pin) connection cable.
The meter must be in measurement mode to communicate.
The meter RS232 port is optoisolated and transmits data with a Baud
rate of 2400 bps.
Retrieve the GLP data and request the meter parameters and the
current reading (of the current range only) from the PC directly. It is
also possible to send a command from the PC to set the meter in a
different range.
To communicate with the PC use the HI 92000 communication soft-
ware. The software is provided with an exhaustive on-line guide of all
the commands available and allows data printing, plotting and ex-
porting.
DATA TRANSFER TO PCDATA TRANSFER TO PC
DATA TRANSFER TO PCDATA TRANSFER TO PC
DATA TRANSFER TO PC