Good Laboratory Practice (GLP) is a set of functions that allows the
storage and retrieval of data regarding the status of the system.
After a successful calibration, the meter automatically stores the date
and time of calibration, the calibration solution used and the resulting
cell constant value. All such information can be recalled by the user.
To view last calibration data select the desired
range (EC or NaCl) and press the GLP key.
The first information appearing on the LCD
is the meter "id" code.
By repeatedly pressing the RANGE key, the GLP
data are displayed in the following order:
Last calibration date.
The following table lists the setup items, their valid range of values
and the factory settings (default):
Item Description Valid values Default
tc Temp. compensation coeff. 0.00 to 6.00 %/C 1.90
tcE Temp. compensation mode Atc, Mtc, notc Atc
tdS TDS factor 0.40 to 0.80 0.50
CEL Cell constant (K) 0.500 to 1.700 1.000
AoF Auto-Off enabled On, OFF OFF
YEA Year 1999 to 2098 1999
dAt Date (DD.MM) 01.01 to 31.12 01.01
hou Time (hh.mm) 00.00 to 23.59 00.00
id Meter identification code 0000 to 9999 0000
vEr Firmware release
Notes:
I Once enabled, the Auto-Off time is fixed at 5 minutes.
II Assigning an ID code is helpful in identifying a particular meter
from others.